It is mainly used to calibrate non-contact temperature sensors. A step change in temperature is achievable within seconds. The graphite tube has an emitter diameter of 25mm (1.00″) and an emissivity of +0.99. The target temperature is sensed by a non-contact temperature sensor/Pyrometer which gives feedback to the PID controller to regulate the cavity temperature precisely to the desired set points. The remote set point is achieved via the RS-232 C communication port. All safety features to protect equipment and operator are easily accessible.
Safety Interlock Features
The system will switch off in case of:
- Water flow failure
- Increase in water temperature
- Argon gas flow failure
- Over Current
Model Details
Model | Fastcal 3000 |
---|---|
Download | Datasheet |
Temperature Range | 600°C to 3000°C |
Temperature Resolution | 0.1°C |
Purge Gas | Argon gas flow with 7-10 LPM respective |
Controlling Sensor | Pyrometer |
Method of Control | PID controller Eurotherm make 3504 (Option – HMI & Data Logging) |
Emissivity | 0.99 |
Heating Time | Approx 5 Minute from 500 to 2700°C |
Ambient Temperature | 10°C to 50°C |
Power Requirement | 440V,Single Phase AC 50/60 Hz 50 KW |
Dimensions | 1700mm(H) x 900mm(W) x 1200mm(D) |
Cavity | Graphite Dual cavity blackbody, one cavity is control and one cavity for measurement |
Heating Aperture | 25mm |
Weight | 300 Kg |